*** MIDAS logfile (05SEPpl1.2) *** Fr, 29 Sep 2006 19:05:54 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 29 Sep 2006 19:06:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060929 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.27 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060929 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1712.6 1712.8 878.4 0.11 0.10 1.95 2 1.09 3409.4 3412.1 1739.0 -0.15 -0.11 1.96 3 2.17 6836.4 6837.6 3457.3 0.09 0.11 1.98 4 3.25 10232.3 10232.4 5095.5 -0.08 -0.07 2.01 5 4.88 15364.4 15363.7 7544.1 0.04 0.03 2.04 6 6.50 20467.9 20464.6 10302.6 -0.05 -0.07 1.99 7 8.13 25604.9 25604.1 13135.9 0.03 0.02 1.95 8 10.84 34123.6 34124.0 16841.7 -0.00 -0.01 2.03 9 13.55 42651.0 42653.0 20813.2 -0.01 -0.01 2.05 10 16.25 51189.0 51193.1 25542.0 0.01 0.01 2.00 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3464 to 103544 e- Average count rate measured: 3149 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.02 +/- 0.02 e-/ADU RON = 4.00 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay