*** MIDAS logfile (05SEPpl1.2) *** We, 13 Sep 2006 14:35:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 13 Sep 2006 14:35:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060913 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.16 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060913 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1717.8 1718.1 903.0 0.04 0.03 1.90 2 1.09 3416.5 3419.5 1697.9 -0.22 -0.17 2.01 3 2.17 6852.4 6855.4 3438.5 0.11 0.13 1.99 4 3.25 10255.6 10255.6 5197.0 -0.07 -0.08 1.97 5 4.88 15396.0 15399.0 7807.5 0.05 0.05 1.97 6 6.50 20514.5 20510.6 10448.5 -0.03 -0.05 1.96 7 8.13 25662.7 25664.4 12856.2 0.06 0.06 2.00 8 10.84 34190.6 34193.0 16998.7 -0.01 -0.01 2.01 9 13.55 42766.0 42761.9 21245.7 0.06 0.05 2.01 10 16.25 51285.9 51285.0 25628.9 0.00 -0.01 2.00 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3454 to 103112 e- Average count rate measured: 3155 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.01 e-/ADU RON = 4.18 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay