*** MIDAS logfile (05SEPpl1.2) *** We, 06 Sep 2006 13:57:25 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Sep 2006 13:57:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060906 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.40 stddev = 1.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060906 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1713.9 1714.8 855.9 0.09 0.13 2.00 2 1.09 3414.5 3413.0 1718.7 -0.14 -0.18 1.99 3 2.17 6843.6 6841.3 3496.2 0.10 0.08 1.96 4 3.25 10243.7 10245.6 5147.3 -0.07 -0.06 1.99 5 4.88 15379.7 15378.5 7607.9 0.03 0.03 2.02 6 6.50 20488.7 20490.8 10436.9 -0.06 -0.05 1.96 7 8.13 25635.1 25636.9 12533.9 0.04 0.05 2.05 8 10.84 34163.1 34160.8 17289.0 0.00 -0.00 1.98 9 13.55 42708.7 42708.9 20991.9 0.01 0.01 2.03 10 16.25 51242.4 51240.6 26066.3 0.00 -0.00 1.97 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3413 to 102027 e- Average count rate measured: 3153 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.99 +/- 0.02 e-/ADU RON = 3.96 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay