*** MIDAS logfile (05SEPpl1.2) *** Sa, 26 Aug 2006 19:18:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 26 Aug 2006 19:18:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060826 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.45 stddev = 2.19 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060826 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1698.1 1698.5 847.4 0.07 0.08 2.00 2 1.09 3382.6 3383.3 1629.9 -0.16 -0.14 2.08 3 2.17 6779.3 6780.7 3511.4 0.09 0.08 1.93 4 3.25 10150.8 10151.2 4983.2 -0.07 -0.07 2.04 5 4.88 15237.8 15240.4 7690.0 0.03 0.04 1.98 6 6.50 20300.7 20303.2 10048.4 -0.06 -0.05 2.02 7 8.13 25401.5 25402.2 12893.9 0.05 0.04 1.97 8 10.84 33853.0 33848.6 16930.4 0.01 -0.01 2.00 9 13.55 42315.0 42315.3 20727.9 0.01 0.00 2.04 10 16.25 50789.5 50790.3 24756.0 0.03 0.03 2.05 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3470 to 103785 e- Average count rate measured: 3124 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.04 +/- 0.02 e-/ADU RON = 4.47 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay