*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 14:41:19 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 14:41:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060820 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.50 stddev = 2.19 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1716.2 1716.3 861.4 0.05 0.04 1.99 2 1.09 3412.0 3412.5 1759.1 -0.20 -0.21 1.94 3 2.17 6838.5 6840.3 3381.3 0.07 0.09 2.02 4 3.25 10235.5 10236.1 5175.0 -0.09 -0.08 1.98 5 4.88 15370.6 15372.1 7869.0 0.05 0.06 1.95 6 6.50 20477.9 20474.8 10218.0 -0.03 -0.05 2.00 7 8.13 25609.8 25617.5 12890.8 0.04 0.06 1.99 8 10.84 34135.5 34135.9 16983.0 0.01 0.01 2.01 9 13.55 42688.1 42682.5 21067.6 0.06 0.04 2.03 10 16.25 51212.4 51209.0 25456.7 0.04 0.03 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3468 to 103473 e- Average count rate measured: 3150 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.02 +/- 0.01 e-/ADU RON = 4.42 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay