*** MIDAS logfile (05SEPpl1.2) *** Th, 20 Jul 2006 15:55:35 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 20 Jul 2006 15:55:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060720 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.54 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1582.4 1580.6 790.8 0.04 -0.03 2.00 2 1.54 3154.0 3154.0 1607.3 0.01 0.02 1.96 3 3.08 6296.2 6298.3 3267.7 -0.05 -0.03 1.93 4 4.62 9441.9 9443.8 4681.2 -0.02 -0.01 2.02 5 6.93 14145.7 14144.6 7172.2 -0.12 -0.12 1.97 6 9.23 18872.6 18874.3 9273.9 -0.05 -0.04 2.04 7 11.54 23602.8 23601.9 11718.0 0.02 0.02 2.01 8 15.39 31471.4 31470.4 15597.0 0.02 0.02 2.02 9 19.23 39355.6 39358.4 19441.3 0.07 0.08 2.02 10 23.08 47231.0 47228.9 23742.2 0.09 0.09 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3178 to 94911 e- Average count rate measured: 2044 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.01 e-/ADU RON = 4.07 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay