*** MIDAS logfile (05SEPpl1.2) *** Mo, 19 Jun 2006 07:34:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 19 Jun 2006 07:34:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060619 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.12 stddev = 2.10 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060619 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1604.6 1603.8 792.6 0.05 0.04 2.02 2 1.54 3196.6 3197.0 1574.0 -0.02 0.00 2.03 3 3.08 6383.9 6386.7 3249.3 -0.05 -0.02 1.96 4 4.62 9571.9 9573.9 4747.5 -0.03 -0.01 2.02 5 6.93 14346.1 14343.4 7227.0 -0.10 -0.12 1.99 6 9.23 19134.4 19132.1 9651.5 -0.05 -0.06 1.98 7 11.54 23935.4 23933.9 12203.7 0.04 0.03 1.96 8 15.39 31902.0 31903.5 15808.2 0.01 0.01 2.02 9 19.23 39908.9 39906.4 19597.2 0.09 0.08 2.04 10 23.08 47872.1 47867.7 22852.7 0.06 0.05 2.09 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3321 to 99086 e- Average count rate measured: 2073 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.07 +/- 0.02 e-/ADU RON = 4.34 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay