*** MIDAS logfile (05SEPpl1.2) *** Sa, 06 May 2006 14:57:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 06 May 2006 14:57:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060506 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.00 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060506 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1648.4 1647.9 834.1 0.05 0.02 1.98 2 1.54 3281.3 3281.1 1631.3 -0.11 -0.11 2.01 3 3.08 6557.6 6560.4 3226.4 -0.04 -0.01 2.03 4 4.62 9830.7 9832.3 4770.6 -0.03 -0.02 2.06 5 6.93 14733.5 14731.7 7327.2 -0.09 -0.10 2.01 6 9.23 19652.2 19650.6 9718.7 -0.03 -0.04 2.02 7 11.54 24583.0 24583.1 12175.1 0.06 0.06 2.02 8 15.39 32763.6 32757.1 16690.2 0.02 0.01 1.96 9 19.23 40996.1 40993.3 21216.3 0.13 0.13 1.93 10 23.08 49145.9 49144.1 24171.3 0.04 0.04 2.03 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3271 to 97545 e- Average count rate measured: 2128 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.98 +/- 0.03 e-/ADU RON = 4.05 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay