*** MIDAS logfile (05SEPpl1.2) *** Fr, 07 Apr 2006 13:11:49 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 07 Apr 2006 13:12:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060407 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 283.50 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060407 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1667.8 1666.2 829.3 0.08 0.04 2.01 2 1.54 3320.5 3318.6 1663.7 -0.01 -0.04 2.00 3 3.08 6627.8 6628.2 3367.0 -0.05 -0.04 1.97 4 4.62 9937.4 9938.6 4993.2 -0.02 -0.01 1.99 5 6.93 14889.4 14887.0 7608.6 -0.11 -0.12 1.96 6 9.23 19863.8 19859.7 10035.7 -0.04 -0.05 1.98 7 11.54 24841.0 24842.6 12850.2 0.02 0.04 1.93 8 15.39 33114.6 33115.2 16506.2 0.01 0.02 2.01 9 19.23 41417.7 41419.7 20506.8 0.08 0.09 2.02 10 23.08 49688.0 49686.7 24282.8 0.06 0.06 2.05 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3393 to 101136 e- Average count rate measured: 2151 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.04 +/- 0.02 e-/ADU RON = 4.08 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay