*** MIDAS logfile (05SEPpl1.2) *** Sa, 11 Mar 2006 18:31:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 11 Mar 2006 18:32:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060311 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 286.18 stddev = 2.11 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060311 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1678.4 1680.9 848.1 0.03 0.01 1.98 2 1.54 3345.6 3347.5 1654.7 -0.05 -0.07 2.02 3 3.08 6686.1 6687.8 3315.7 -0.03 -0.03 2.02 4 4.62 10025.0 10026.6 5048.0 -0.01 -0.01 1.99 5 6.93 15018.4 15019.3 7399.1 -0.12 -0.11 2.03 6 9.23 20037.6 20038.8 10232.6 -0.04 -0.03 1.96 7 11.54 25066.8 25063.4 12785.7 0.05 0.05 1.96 8 15.39 33407.4 33404.9 16651.3 0.01 0.02 2.01 9 19.23 41793.5 41791.6 20710.3 0.10 0.11 2.02 10 23.08 50123.4 50128.4 24932.0 0.05 0.07 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3377 to 100790 e- Average count rate measured: 2170 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.01 e-/ADU RON = 4.24 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay