*** MIDAS logfile (05SEPpl1.2) *** Sa, 25 Feb 2006 12:18:35 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 25 Feb 2006 12:18:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060225 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.88 stddev = 2.01 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060225 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1687.1 1688.0 830.9 0.06 0.02 2.03 2 1.54 3362.2 3362.8 1658.6 -0.02 -0.07 2.03 3 3.08 6715.3 6717.5 3389.7 -0.04 -0.03 1.98 4 4.62 10071.0 10070.5 4952.6 -0.01 -0.01 2.03 5 6.93 15086.2 15084.9 7625.1 -0.12 -0.11 1.98 6 9.23 20129.1 20124.5 10213.6 -0.04 -0.04 1.97 7 11.54 25176.1 25173.8 12505.2 0.04 0.05 2.01 8 15.39 33555.6 33558.8 17070.1 0.00 0.03 1.97 9 19.23 41976.9 41975.7 20619.7 0.09 0.11 2.04 10 23.08 50340.7 50343.6 24948.5 0.03 0.06 2.02 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3404 to 101531 e- Average count rate measured: 2180 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.02 +/- 0.01 e-/ADU RON = 4.05 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay