*** MIDAS logfile (05SEPpl1.2) *** Fr, 03 Feb 2006 17:33:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 03 Feb 2006 17:33:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060203 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 286.10 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060203 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1704.2 1704.7 862.9 0.03 0.05 1.97 2 1.54 3398.1 3395.8 1719.1 0.00 -0.04 1.98 3 3.08 6783.5 6784.0 3359.6 -0.03 -0.02 2.02 4 4.62 10169.1 10168.7 5108.8 -0.02 -0.02 1.99 5 6.93 15233.3 15237.1 7654.7 -0.12 -0.10 1.99 6 9.23 20327.1 20325.3 10147.2 -0.03 -0.03 2.00 7 11.54 25416.9 25420.6 12558.1 0.02 0.04 2.02 8 15.39 33882.3 33884.3 16742.0 0.01 0.01 2.02 9 19.23 42385.2 42378.0 21356.7 0.09 0.07 1.98 10 23.08 50839.0 50834.8 25211.9 0.05 0.04 2.02 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3426 to 102185 e- Average count rate measured: 2201 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.01 e-/ADU RON = 4.16 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay