*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 14:41:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 14:42:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060114 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.93 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1708.3 1706.8 854.5 -0.00 -0.04 2.00 2 1.54 3403.7 3405.5 1759.2 -0.00 0.01 1.93 3 3.08 6795.1 6794.1 3429.9 -0.06 -0.06 1.98 4 4.62 10189.6 10189.4 5130.3 -0.01 -0.01 1.99 5 6.93 15266.4 15266.7 7644.6 -0.10 -0.10 2.00 6 9.23 20364.7 20367.0 10260.3 -0.04 -0.03 1.98 7 11.54 25465.8 25467.2 12729.8 0.02 0.02 2.00 8 15.39 33960.4 33963.2 16903.3 0.04 0.05 2.01 9 19.23 42452.8 42455.7 21089.7 0.05 0.06 2.01 10 23.08 50963.6 50964.5 25515.8 0.10 0.10 2.00 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3428 to 102319 e- Average count rate measured: 2206 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.00 e-/ADU RON = 4.08 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay