*** MIDAS logfile (05SEPpl1.2) *** We, 11 Jan 2006 15:03:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 11 Jan 2006 15:03:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 060111 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 286.45 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_060111 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1682.8 1682.8 820.0 -0.05 -0.08 2.05 2 1.54 3374.2 3375.7 1650.8 0.25 0.25 2.04 3 3.08 6725.5 6723.4 3404.7 -0.09 -0.14 1.98 4 4.62 10100.9 10104.8 5070.3 0.07 0.08 1.99 5 6.93 15123.7 15125.2 7526.1 -0.12 -0.13 2.01 6 9.23 20187.0 20190.9 10141.8 -0.01 -0.01 1.99 7 11.54 25214.4 25226.2 12745.2 -0.07 -0.05 1.98 8 15.39 33675.6 33682.6 16387.1 0.09 0.09 2.06 9 19.23 41944.5 41997.3 20628.0 -0.26 -0.16 2.03 10 23.08 50555.1 50557.0 24703.1 0.18 0.16 2.05 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3446 to 103515 e- Average count rate measured: 2187 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.05 +/- 0.01 e-/ADU RON = 4.27 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay