*** MIDAS logfile (05SEPpl1.2) *** Mo, 26 Dec 2005 14:01:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 26 Dec 2005 14:01:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051226 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 286.25 stddev = 2.01 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_051226 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1687.1 1688.1 825.4 -0.02 0.03 2.04 2 1.54 3365.6 3368.0 1694.9 -0.00 0.03 1.99 3 3.08 6720.3 6721.6 3437.7 -0.06 -0.06 1.95 4 4.62 10082.1 10081.3 5060.8 0.01 -0.02 1.99 5 6.93 15099.7 15101.5 7422.8 -0.12 -0.13 2.03 6 9.23 20150.0 20150.0 10285.6 -0.02 -0.04 1.96 7 11.54 25197.2 25196.8 12442.7 0.03 0.02 2.03 8 15.39 33599.6 33599.0 17021.9 0.04 0.03 1.97 9 19.23 42002.7 42005.8 21067.6 0.05 0.05 1.99 10 23.08 50422.2 50424.6 25186.7 0.10 0.09 2.00 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3369 to 100669 e- Average count rate measured: 2183 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.00 +/- 0.01 e-/ADU RON = 4.01 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay