*** MIDAS logfile (05SEPpl1.1) *** Sa, 26 Nov 2005 15:22:50 Page 1 *** MIDAS logfile (vers. 05SEPpl1.1) *** Sa, 26 Nov 2005 15:23:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051126 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 286.79 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_051126 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1678.6 1679.2 849.7 0.06 0.06 1.98 2 1.54 3346.4 3349.2 1635.8 -0.01 0.03 2.05 3 3.08 6684.9 6685.2 3285.8 -0.03 -0.05 2.03 4 4.62 10022.6 10021.6 4943.6 -0.01 -0.04 2.03 5 6.93 15014.9 15016.3 9257.7 -0.12 -0.12 1.62 6 9.23 20033.6 20034.4 10024.5 -0.04 -0.04 2.00 7 11.54 25052.0 25054.6 12568.3 0.02 0.01 1.99 8 15.39 33404.4 33404.7 16796.9 0.02 0.01 1.99 9 19.23 41759.5 41773.6 20236.2 0.04 0.06 2.06 10 23.08 50122.8 50124.0 24925.3 0.07 0.06 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3424 to 102233 e- Average count rate measured: 2170 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.04 +/- 0.05 e-/ADU RON = 4.16 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay