*** MIDAS logfile (03SEPpl1.2) *** Sa, 08 Oct 2005 11:31:18 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 08 Oct 2005 11:31:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 051008 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 286.23 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_051008 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1696.8 1698.5 851.6 0.01 0.04 1.99 2 1.54 3379.4 3379.0 1715.8 -0.10 -0.10 1.97 3 3.08 6753.7 6753.4 3334.1 -0.02 -0.03 2.03 4 4.62 10123.7 10123.6 5205.2 -0.02 -0.02 1.94 5 6.93 15167.6 15167.8 7382.7 -0.11 -0.11 2.05 6 9.23 20233.2 20235.8 10249.7 -0.04 -0.03 1.97 7 11.54 25316.9 25317.0 12501.5 0.07 0.07 2.03 8 15.39 33735.5 33729.4 17139.5 0.02 0.00 1.97 9 19.23 42220.5 42217.8 21425.0 0.15 0.14 1.97 10 23.08 50603.5 50605.4 25099.6 0.03 0.04 2.02 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3387 to 100961 e- Average count rate measured: 2192 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.00 +/- 0.02 e-/ADU RON = 4.00 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay