*** MIDAS logfile (03SEPpl1.2) *** Fr, 05 Nov 2004 12:33:32 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 05 Nov 2004 12:33:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041105 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.95 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_041105 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1644.2 1646.1 852.5 0.03 0.09 1.93 2 1.54 3274.3 3273.3 1691.7 -0.02 -0.05 1.94 3 3.08 6532.5 6534.2 3247.1 -0.06 -0.05 2.01 4 4.62 9792.8 9794.9 4998.5 -0.04 -0.03 1.96 5 6.93 14672.9 14671.5 7366.0 -0.11 -0.13 1.99 6 9.23 19574.8 19573.9 9799.7 -0.04 -0.05 2.00 7 11.54 24483.2 24485.1 12346.4 0.04 0.04 1.98 8 15.39 32638.9 32637.6 16188.2 0.03 0.02 2.02 9 19.23 40815.2 40819.0 20307.4 0.08 0.08 2.01 10 23.08 48975.4 48974.3 23145.6 0.08 0.07 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3417 to 101730 e- Average count rate measured: 2120 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.08 +/- 0.03 e-/ADU RON = 4.28 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay