*** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Jul 2004 12:45:41 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Jul 2004 12:45:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040720 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.62 stddev = 2.15 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_040720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1676.2 1676.0 856.8 0.09 0.07 1.96 2 1.54 3329.8 3327.3 1728.4 -0.11 -0.14 1.93 3 3.08 6649.3 6647.1 3387.0 -0.04 -0.05 1.96 4 4.62 9964.7 9963.9 5108.3 -0.03 -0.03 1.95 5 6.93 14930.4 14929.6 7414.2 -0.11 -0.10 2.01 6 9.23 19917.3 19913.1 10058.9 -0.03 -0.03 1.98 7 11.54 24911.7 24912.0 12566.6 0.05 0.07 1.98 8 15.39 33197.2 33198.2 17143.8 0.01 0.02 1.94 9 19.23 41540.4 41534.0 20415.2 0.12 0.12 2.03 10 23.08 49809.8 49804.4 24852.2 0.05 0.06 2.00 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3366 to 100021 e- Average count rate measured: 2156 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.02 e-/ADU RON = 4.33 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay