*** MIDAS logfile (vers. 02SEPpl1.3) *** Sa, 03 Jul 2004 16:50:05 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Sa, 03 Jul 2004 16:50:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040703 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.82 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_040703 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1671.8 1672.4 875.8 0.08 0.07 1.91 2 1.54 3325.7 3325.3 1685.3 -0.05 -0.06 1.97 3 3.08 6639.0 6639.4 3326.1 -0.07 -0.07 2.00 4 4.62 9954.5 9954.9 5040.9 -0.03 -0.02 1.97 5 6.93 14914.0 14912.2 7489.9 -0.12 -0.12 1.99 6 9.23 19897.9 19895.8 9888.6 -0.03 -0.04 2.01 7 11.54 24882.1 24883.0 12714.1 0.03 0.03 1.96 8 15.39 33173.6 33175.4 16762.5 0.03 0.04 1.98 9 19.23 41484.0 41485.4 20570.1 0.08 0.08 2.02 10 23.08 49777.5 49778.7 25562.9 0.08 0.08 1.95 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3298 to 98168 e- Average count rate measured: 2155 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.97 +/- 0.02 e-/ADU RON = 4.06 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay