*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 08 Feb 2004 13:24:57 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 08 Feb 2004 13:25:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040208 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.27 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_040208 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1772.1 1772.5 919.0 -0.02 -0.01 1.93 2 1.54 3530.9 3530.2 1769.4 -0.01 -0.03 2.00 3 3.08 7046.1 7046.7 3609.6 -0.07 -0.06 1.95 4 4.62 10570.2 10568.7 5399.4 0.02 0.01 1.96 5 6.93 15826.4 15826.4 7867.1 -0.12 -0.12 2.01 6 9.23 21119.9 21119.8 10691.1 -0.02 -0.02 1.98 7 11.54 26404.2 26407.4 13142.0 0.03 0.03 2.01 8 15.39 35215.6 35217.3 17439.9 0.06 0.06 2.02 9 19.23 44001.6 44007.7 22141.2 0.03 0.04 1.99 10 23.08 52835.0 52836.7 26293.0 0.10 0.10 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3561 to 106165 e- Average count rate measured: 2287 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.01 e-/ADU RON = 4.11 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay