*** MIDAS logfile (05SEPpl1.2) *** Sa, 01 Mar 2008 15:16:14 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 01 Mar 2008 15:16:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 080301 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.59 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_080301 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1604.4 1604.3 758.6 0.15 0.14 2.11 2 4.34 3196.5 3197.2 1511.0 -0.22 -0.20 2.12 3 8.67 6399.5 6399.8 3053.7 -0.10 -0.10 2.10 4 13.00 9597.5 9597.8 4617.1 -0.12 -0.12 2.08 5 19.51 14401.8 14402.1 7018.4 -0.09 -0.09 2.05 6 26.01 19211.8 19212.0 9247.3 -0.04 -0.03 2.08 7 32.51 24025.2 24025.0 11698.3 0.01 0.01 2.05 8 43.34 32064.3 32063.9 15289.5 0.11 0.10 2.10 9 54.17 40087.6 40090.4 19017.8 0.12 0.13 2.11 10 65.01 48127.7 48127.4 22624.4 0.17 0.17 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3399 to 101965 e- Average count rate measured: 739 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.19 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay