*** MIDAS logfile (05SEPpl1.2) *** Sa, 09 Feb 2008 14:11:13 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 09 Feb 2008 14:11:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 080209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.90 stddev = 1.99 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_080209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1614.5 1614.0 767.7 0.14 0.15 2.10 2 4.34 3217.4 3217.6 1538.1 -0.21 -0.19 2.09 3 8.67 6440.9 6440.0 3077.5 -0.10 -0.11 2.09 4 13.00 9658.9 9659.8 4619.7 -0.13 -0.12 2.09 5 19.51 14495.4 14495.3 7111.2 -0.08 -0.09 2.04 6 26.01 19335.6 19336.0 9428.5 -0.03 -0.04 2.05 7 32.51 24180.5 24181.4 11623.7 0.01 0.01 2.08 8 43.34 32271.1 32271.4 15494.1 0.10 0.10 2.08 9 54.17 40348.7 40353.6 19149.3 0.13 0.13 2.11 10 65.01 48436.9 48438.5 22896.5 0.17 0.16 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3408 to 102259 e- Average count rate measured: 744 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.01 e-/ADU RON = 4.20 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay