*** MIDAS logfile (05SEPpl1.2) *** We, 30 Jan 2008 16:13:02 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 30 Jan 2008 16:13:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 080130 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.93 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_080130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1614.1 1613.7 774.0 0.16 0.14 2.09 2 4.34 3215.3 3215.3 1545.8 -0.22 -0.21 2.08 3 8.67 6437.4 6436.8 3105.3 -0.09 -0.10 2.07 4 13.00 9651.8 9652.5 4622.9 -0.14 -0.13 2.09 5 19.51 14486.1 14485.2 6985.5 -0.08 -0.09 2.07 6 26.01 19323.7 19322.6 9317.1 -0.03 -0.04 2.07 7 32.51 24166.7 24166.4 11713.5 0.02 0.02 2.06 8 43.34 32245.5 32245.8 15457.5 0.09 0.09 2.09 9 54.17 40333.1 40334.4 18983.6 0.15 0.16 2.12 10 65.01 48397.1 48397.2 22935.9 0.15 0.15 2.11 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3412 to 102306 e- Average count rate measured: 743 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.01 e-/ADU RON = 4.19 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay