*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Jan 2008 14:33:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Jan 2008 14:33:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 080110 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.68 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_080110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1621.1 1621.3 768.8 0.12 0.13 2.11 2 4.34 3235.3 3235.5 1523.9 -0.13 -0.13 2.12 3 8.67 6471.6 6470.8 3095.2 -0.13 -0.15 2.09 4 13.00 9711.4 9711.8 4667.6 -0.10 -0.10 2.08 5 19.51 14569.7 14570.2 7102.5 -0.09 -0.09 2.05 6 26.01 19436.5 19436.9 9278.1 -0.03 -0.04 2.09 7 32.51 24294.7 24304.6 11776.6 -0.04 -0.01 2.06 8 43.34 32444.8 32448.7 15700.8 0.12 0.12 2.07 9 54.17 40530.4 40534.4 19190.4 0.06 0.06 2.11 10 65.01 48709.2 48713.6 22713.5 0.21 0.21 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3445 to 103524 e- Average count rate measured: 748 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.20 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.21/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay