*** MIDAS logfile (05SEPpl1.2) *** Th, 27 Dec 2007 15:49:37 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 27 Dec 2007 15:49:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071227 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.66 stddev = 1.95 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071227 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1614.1 1614.3 762.5 0.12 0.13 2.12 2 4.34 3218.4 3219.1 1538.3 -0.20 -0.19 2.09 3 8.67 6443.5 6443.5 3086.2 -0.10 -0.10 2.09 4 13.00 9664.5 9663.2 4651.1 -0.11 -0.13 2.08 5 19.51 14501.9 14502.2 6953.4 -0.08 -0.08 2.09 6 26.01 19345.1 19345.4 9312.2 -0.03 -0.03 2.08 7 32.51 24191.8 24190.8 11593.2 0.01 0.00 2.09 8 43.34 32284.6 32285.2 15503.2 0.10 0.10 2.08 9 54.17 40366.6 40370.6 19148.3 0.13 0.13 2.11 10 65.01 48458.0 48459.9 22840.7 0.17 0.17 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3413 to 102443 e- Average count rate measured: 744 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.01 e-/ADU RON = 4.12 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay