*** MIDAS logfile (05SEPpl1.2) *** Th, 20 Dec 2007 16:50:10 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 20 Dec 2007 16:50:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.19 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1612.7 1613.2 761.7 0.11 0.14 2.12 2 4.34 3217.1 3217.2 1534.8 -0.17 -0.18 2.10 3 8.67 6437.7 6437.9 3076.4 -0.11 -0.11 2.09 4 13.00 9657.1 9658.2 4651.5 -0.11 -0.11 2.08 5 19.51 14490.0 14490.6 6977.3 -0.08 -0.09 2.08 6 26.01 19330.7 19329.6 9307.0 -0.03 -0.04 2.08 7 32.51 24169.7 24170.5 11840.4 -0.00 -0.00 2.04 8 43.34 32261.2 32262.6 15552.5 0.11 0.10 2.07 9 54.17 40325.4 40329.4 19447.8 0.10 0.11 2.07 10 65.01 48423.0 48426.5 22521.4 0.17 0.17 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3411 to 102406 e- Average count rate measured: 744 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.02 e-/ADU RON = 4.17 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay