*** MIDAS logfile (05SEPpl1.2) *** Su, 09 Dec 2007 14:54:15 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 09 Dec 2007 14:54:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.65 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1605.6 1605.7 759.5 0.15 0.16 2.11 2 4.34 3200.7 3200.2 1499.9 -0.20 -0.21 2.13 3 8.67 6407.2 6406.9 3047.5 -0.11 -0.11 2.10 4 13.00 9610.3 9609.9 4620.5 -0.12 -0.12 2.08 5 19.51 14420.8 14420.2 6965.6 -0.08 -0.09 2.07 6 26.01 19235.5 19237.4 9413.9 -0.04 -0.03 2.04 7 32.51 24056.8 24055.9 11581.0 0.01 0.01 2.08 8 43.34 32104.9 32106.6 15408.1 0.10 0.10 2.08 9 54.17 40141.2 40141.0 18963.8 0.12 0.12 2.12 10 65.01 48188.0 48189.7 22180.2 0.16 0.17 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3445 to 103397 e- Average count rate measured: 740 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.22 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay