*** MIDAS logfile (05SEPpl1.2) *** Th, 22 Nov 2007 14:39:09 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 22 Nov 2007 14:39:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071122 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.12 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071122 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1610.2 1610.1 761.9 0.14 0.13 2.11 2 4.34 3211.3 3212.2 1519.6 -0.16 -0.14 2.11 3 8.67 6426.4 6425.6 3020.7 -0.11 -0.12 2.13 4 13.00 9639.7 9640.8 4637.5 -0.11 -0.11 2.08 5 19.50 14464.3 14464.0 6946.4 -0.08 -0.09 2.08 6 26.00 19295.1 19295.6 9330.6 -0.04 -0.04 2.07 7 32.51 24126.1 24126.8 11672.3 -0.01 -0.01 2.07 8 43.34 32205.6 32207.2 15563.2 0.11 0.11 2.07 9 54.17 40248.8 40255.5 18921.9 0.09 0.10 2.13 10 65.01 48343.0 48350.2 22241.0 0.18 0.19 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3455 to 103755 e- Average count rate measured: 742 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.24 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay