*** MIDAS logfile (05SEPpl1.2) *** We, 14 Nov 2007 16:43:05 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 14 Nov 2007 16:43:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.42 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1598.3 1597.8 760.3 0.16 0.15 2.10 2 4.34 3184.3 3185.4 1509.2 -0.22 -0.20 2.11 3 8.67 6377.2 6377.0 3083.8 -0.09 -0.09 2.07 4 13.00 9563.3 9562.4 4611.3 -0.12 -0.13 2.07 5 19.51 14351.9 14350.7 7054.5 -0.08 -0.08 2.03 6 26.01 19143.0 19142.2 9301.8 -0.04 -0.04 2.06 7 32.51 23942.5 23943.4 11626.9 0.02 0.02 2.06 8 43.34 31945.5 31941.9 15297.6 0.09 0.08 2.09 9 54.17 39954.3 39953.3 18860.2 0.15 0.14 2.12 10 65.01 47936.7 47940.3 22315.3 0.13 0.14 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3413 to 102382 e- Average count rate measured: 736 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.22 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay