*** MIDAS logfile (05SEPpl1.2) *** Sa, 03 Nov 2007 15:04:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 03 Nov 2007 15:05:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.40 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1610.6 1611.1 755.8 0.15 0.13 2.13 2 4.34 3211.0 3210.5 1546.6 -0.19 -0.21 2.08 3 8.67 6427.7 6428.1 3050.7 -0.10 -0.09 2.11 4 13.00 9640.1 9639.5 4639.6 -0.12 -0.12 2.08 5 19.51 14465.5 14465.9 7002.7 -0.09 -0.08 2.07 6 26.01 19297.4 19295.0 9417.2 -0.04 -0.04 2.05 7 32.51 24131.0 24131.2 11662.2 0.00 0.01 2.07 8 43.34 32204.2 32204.2 15571.8 0.09 0.10 2.07 9 54.17 40270.0 40269.4 18964.4 0.13 0.14 2.12 10 65.01 48336.6 48337.2 22609.6 0.16 0.17 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3425 to 102762 e- Average count rate measured: 742 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.20 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay