*** MIDAS logfile (05SEPpl1.2) *** Mo, 29 Oct 2007 12:37:59 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 29 Oct 2007 12:38:13 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071029 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.38 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071029 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1613.2 1613.3 762.0 0.13 0.14 2.12 2 4.34 3216.7 3217.0 1535.6 -0.19 -0.19 2.09 3 8.67 6438.7 6439.1 3096.6 -0.11 -0.10 2.08 4 13.00 9657.7 9657.7 4687.0 -0.12 -0.12 2.06 5 19.51 14492.5 14491.3 7033.5 -0.08 -0.09 2.06 6 26.01 19330.7 19331.5 9301.2 -0.04 -0.04 2.08 7 32.51 24174.1 24174.0 11669.1 0.00 0.00 2.07 8 43.34 32265.3 32265.6 15550.8 0.11 0.10 2.07 9 54.17 40337.8 40338.3 19012.9 0.12 0.12 2.12 10 65.01 48427.2 48428.9 22993.3 0.17 0.17 2.11 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3403 to 102162 e- Average count rate measured: 744 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.01 e-/ADU RON = 4.17 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay