*** MIDAS logfile (05SEPpl1.2) *** Fr, 19 Oct 2007 13:55:56 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 19 Oct 2007 13:56:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071019 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.50 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071019 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1592.7 1592.3 759.6 0.16 0.15 2.10 2 4.34 3173.9 3174.3 1502.7 -0.22 -0.20 2.11 3 8.67 6354.2 6354.3 3063.7 -0.10 -0.10 2.07 4 13.00 9530.2 9530.4 4578.3 -0.12 -0.12 2.08 5 19.51 14301.3 14301.5 6996.5 -0.08 -0.08 2.04 6 26.01 19077.1 19077.9 9202.9 -0.04 -0.03 2.07 7 32.51 23857.5 23858.2 11555.6 0.01 0.01 2.06 8 43.34 31837.2 31837.7 15086.5 0.09 0.09 2.11 9 54.17 39817.8 39817.4 18912.0 0.15 0.15 2.11 10 65.01 47785.5 47785.0 22641.6 0.16 0.15 2.11 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3363 to 100922 e- Average count rate measured: 734 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.01 e-/ADU RON = 4.16 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay