*** MIDAS logfile (05SEPpl1.2) *** We, 10 Oct 2007 17:13:50 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 10 Oct 2007 17:14:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071010 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.17 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071010 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1600.5 1600.4 766.6 0.14 0.13 2.09 2 4.34 3191.1 3190.6 1523.2 -0.20 -0.21 2.09 3 8.67 6387.4 6387.6 3067.9 -0.11 -0.10 2.08 4 13.00 9579.9 9579.6 4610.4 -0.12 -0.13 2.08 5 19.51 14376.8 14376.9 6984.1 -0.08 -0.08 2.06 6 26.01 19179.3 19179.3 9325.2 -0.03 -0.03 2.06 7 32.51 23983.8 23984.6 11472.8 0.01 0.01 2.09 8 43.34 32005.4 32007.2 15480.0 0.10 0.10 2.07 9 54.17 40023.1 40022.9 19067.4 0.14 0.14 2.10 10 65.01 48038.4 48039.8 22611.3 0.16 0.16 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3380 to 101442 e- Average count rate measured: 738 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.01 e-/ADU RON = 4.16 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay