*** MIDAS logfile (05SEPpl1.2) *** We, 03 Oct 2007 15:55:42 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 03 Oct 2007 15:55:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071003 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.97 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_071003 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1592.3 1592.9 756.5 0.12 0.16 2.10 2 4.34 3176.1 3174.6 1509.3 -0.17 -0.21 2.10 3 8.67 6356.5 6355.9 3056.0 -0.09 -0.10 2.08 4 13.00 9532.1 9531.9 4608.5 -0.13 -0.13 2.07 5 19.51 14304.7 14305.4 6913.6 -0.08 -0.08 2.07 6 26.01 19082.1 19081.6 9193.6 -0.04 -0.04 2.08 7 32.51 23864.0 23863.4 11535.4 0.01 0.01 2.07 8 43.34 31845.2 31844.8 15286.8 0.09 0.09 2.08 9 54.17 39820.7 39821.8 18594.3 0.13 0.13 2.14 10 65.01 47798.2 47797.4 22139.7 0.16 0.16 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3421 to 102681 e- Average count rate measured: 734 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.22 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay