*** MIDAS logfile (05SEPpl1.2) *** We, 26 Sep 2007 17:45:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 26 Sep 2007 17:45:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070926 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.62 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070926 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1611.8 1612.4 764.3 0.13 0.14 2.11 2 4.34 3214.0 3213.7 1546.8 -0.20 -0.20 2.08 3 8.67 6433.7 6434.0 3061.3 -0.10 -0.10 2.10 4 13.00 9648.8 9649.2 4700.9 -0.13 -0.12 2.05 5 19.51 14480.3 14480.6 7009.9 -0.08 -0.08 2.07 6 26.01 19316.6 19315.8 9354.9 -0.03 -0.04 2.06 7 32.51 24155.5 24156.4 11641.1 0.01 0.01 2.08 8 43.34 32236.2 32236.3 15578.0 0.10 0.10 2.07 9 54.17 40308.6 40310.0 19096.8 0.13 0.13 2.11 10 65.01 48384.1 48386.0 22741.1 0.16 0.16 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3414 to 102454 e- Average count rate measured: 743 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.17 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay