*** MIDAS logfile (05SEPpl1.2) *** We, 19 Sep 2007 18:18:14 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 19 Sep 2007 18:18:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070919 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.58 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070919 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1615.9 1615.6 774.8 0.12 0.14 2.09 2 4.34 3222.1 3221.9 1548.2 -0.19 -0.19 2.08 3 8.67 6450.2 6449.9 3103.2 -0.10 -0.10 2.08 4 13.00 9674.1 9674.8 4706.8 -0.12 -0.12 2.06 5 19.51 14517.3 14516.5 6981.1 -0.08 -0.09 2.08 6 26.01 19365.2 19366.2 9365.1 -0.03 -0.03 2.07 7 32.51 24215.5 24217.2 11881.0 0.00 0.00 2.04 8 43.34 32318.9 32321.3 15513.1 0.10 0.10 2.08 9 54.17 40407.3 40409.4 19063.6 0.12 0.12 2.12 10 65.01 48511.1 48511.5 22400.9 0.17 0.16 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3464 to 104009 e- Average count rate measured: 745 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.22 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay