*** MIDAS logfile (05SEPpl1.2) *** Sa, 15 Sep 2007 17:28:56 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 15 Sep 2007 17:29:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070915 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.47 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070915 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1614.6 1615.1 777.5 0.13 0.14 2.08 2 4.34 3219.8 3220.2 1533.7 -0.18 -0.18 2.10 3 8.67 6445.2 6444.3 3052.2 -0.10 -0.11 2.11 4 13.00 9666.1 9666.1 4656.6 -0.12 -0.12 2.08 5 19.51 14504.0 14504.8 7098.0 -0.09 -0.09 2.04 6 26.01 19349.6 19348.7 9384.3 -0.03 -0.04 2.06 7 32.51 24195.8 24197.5 11590.3 0.00 0.01 2.09 8 43.34 32291.6 32293.3 15690.7 0.10 0.10 2.06 9 54.17 40374.8 40376.3 19503.1 0.12 0.12 2.07 10 65.01 48472.1 48473.7 22530.5 0.17 0.17 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3415 to 102494 e- Average count rate measured: 744 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.02 e-/ADU RON = 4.15 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay