*** MIDAS logfile (05SEPpl1.2) *** We, 29 Aug 2007 17:56:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 29 Aug 2007 17:57:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070829 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.73 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070829 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1617.4 1617.2 765.5 0.16 0.14 2.11 2 4.34 3224.2 3224.5 1530.6 -0.20 -0.20 2.11 3 8.67 6455.2 6455.0 3059.4 -0.09 -0.09 2.11 4 13.00 9678.8 9679.1 4634.4 -0.13 -0.13 2.09 5 19.51 14525.3 14525.7 7053.0 -0.09 -0.09 2.06 6 26.01 19376.3 19377.5 9353.1 -0.04 -0.04 2.07 7 32.51 24231.9 24233.5 11613.4 0.00 0.01 2.09 8 43.34 32339.4 32338.5 15352.2 0.10 0.09 2.11 9 54.17 40436.9 40441.3 18922.8 0.13 0.14 2.14 10 65.01 48545.0 48543.8 22627.1 0.17 0.17 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3464 to 103972 e- Average count rate measured: 746 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 4.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay