*** MIDAS logfile (05SEPpl1.2) *** Mo, 27 Aug 2007 15:09:50 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 27 Aug 2007 15:10:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070827 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.95 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070827 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1613.9 1614.4 765.0 0.16 0.15 2.11 2 4.34 3216.5 3217.4 1546.5 -0.21 -0.21 2.08 3 8.67 6439.2 6439.9 3077.1 -0.11 -0.11 2.09 4 13.00 9657.5 9658.5 4618.4 -0.13 -0.13 2.09 5 19.51 14495.3 14494.7 7149.6 -0.07 -0.08 2.03 6 26.01 19336.4 19334.4 9258.8 -0.03 -0.03 2.09 7 32.51 24182.7 24180.1 11611.2 0.02 0.02 2.08 8 43.34 32264.6 32262.9 15527.8 0.09 0.09 2.08 9 54.17 40349.3 40348.8 18932.2 0.14 0.14 2.13 10 65.01 48423.5 48424.7 22609.5 0.14 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3446 to 103368 e- Average count rate measured: 744 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay