*** MIDAS logfile (05SEPpl1.2) *** Th, 09 Aug 2007 15:53:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 09 Aug 2007 15:53:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070809 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 358.38 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070809 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1627.2 1627.4 776.0 0.14 0.14 2.10 2 4.34 3246.3 3247.0 1546.9 -0.16 -0.15 2.10 3 8.67 6496.0 6496.1 3100.9 -0.12 -0.12 2.09 4 13.00 9745.6 9746.4 4704.3 -0.11 -0.11 2.07 5 19.51 14621.9 14622.0 7015.6 -0.09 -0.10 2.08 6 26.01 19507.4 19508.5 9400.0 -0.04 -0.04 2.08 7 32.51 24390.9 24391.9 11906.8 -0.01 -0.02 2.05 8 43.34 32560.0 32565.5 15415.9 0.11 0.12 2.11 9 54.17 40692.3 40697.7 19222.6 0.09 0.09 2.12 10 65.01 48880.5 48886.7 22840.9 0.19 0.20 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3469 to 104204 e- Average count rate measured: 751 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay