*** MIDAS logfile (05SEPpl1.2) *** Th, 02 Aug 2007 16:15:27 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 02 Aug 2007 16:15:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070802 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 358.18 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070802 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1626.0 1626.7 772.7 0.13 0.15 2.10 2 4.34 3241.8 3241.6 1528.3 -0.20 -0.21 2.12 3 8.67 6490.4 6490.6 3085.1 -0.09 -0.09 2.10 4 13.00 9733.7 9733.4 4639.5 -0.12 -0.12 2.10 5 19.51 14606.2 14607.4 7077.0 -0.08 -0.08 2.06 6 26.01 19485.2 19484.4 9569.4 -0.03 -0.04 2.04 7 32.51 24368.0 24368.7 11779.9 0.02 0.01 2.07 8 43.34 32515.7 32514.5 15737.7 0.09 0.09 2.07 9 54.17 40660.7 40666.8 19391.5 0.13 0.14 2.10 10 65.01 48800.1 48802.7 22857.3 0.15 0.15 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3437 to 103125 e- Average count rate measured: 750 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.02 e-/ADU RON = 4.15 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay