*** MIDAS logfile (05SEPpl1.2) *** We, 25 Jul 2007 15:21:50 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 25 Jul 2007 15:22:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070725 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 358.34 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070725 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1625.1 1624.8 769.7 11.21 25.04 2.11 2 4.34 3235.6 -0.0 937.1 10.79 -100.00 3.45 3 8.67 6482.3 6484.0 3095.4 11.04 24.90 2.09 4 13.00 9718.2 9717.0 4596.3 10.98 24.80 2.11 5 19.51 14588.9 14588.4 6976.1 11.08 24.92 2.09 6 26.01 19458.9 19457.4 9352.9 11.13 24.97 2.08 7 32.51 24340.4 24339.4 11792.6 11.20 25.06 2.06 8 43.34 32459.9 32458.3 15541.0 11.23 25.09 2.09 9 54.17 40619.0 40617.9 19131.4 11.35 25.23 2.12 10 65.01 0.2 0.0 4.0 -100.00 -100.00 0.04 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 0 to 85794 e- Average count rate measured: 636 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.03 e-/ADU RON = 4.15 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = +18.29/-100.00 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay