*** MIDAS logfile (05SEPpl1.2) *** Fr, 20 Jul 2007 13:58:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 20 Jul 2007 13:58:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070720 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 358.03 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1616.8 1616.6 763.6 0.15 0.15 2.12 2 4.34 3222.6 3221.7 1556.1 -0.20 -0.22 2.07 3 8.67 6451.4 6451.0 3069.8 -0.10 -0.10 2.10 4 13.00 9675.5 9673.9 4646.4 -0.12 -0.13 2.08 5 19.51 14518.3 14517.9 7062.8 -0.09 -0.08 2.06 6 26.01 19367.5 19368.3 9325.8 -0.04 -0.03 2.08 7 32.51 24220.8 24220.2 11784.4 0.01 0.01 2.06 8 43.34 32322.3 32322.2 15553.5 0.10 0.10 2.08 9 54.17 40416.7 40416.9 19321.6 0.13 0.14 2.09 10 65.01 48512.1 48513.0 23097.0 0.16 0.16 2.10 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3386 to 101608 e- Average count rate measured: 745 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.09 +/- 0.01 e-/ADU RON = 4.11 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay