*** MIDAS logfile (05SEPpl1.2) *** Sa, 07 Jul 2007 17:31:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 07 Jul 2007 17:31:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070707 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.74 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070707 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1636.8 1636.9 782.7 0.09 0.08 2.09 2 4.34 3270.7 3270.2 1540.0 -0.08 -0.09 2.12 3 8.67 6537.4 6538.3 3089.1 -0.15 -0.15 2.12 4 13.00 9813.3 9814.0 4739.9 -0.09 -0.10 2.07 5 19.51 14719.6 14722.3 7151.1 -0.11 -0.10 2.06 6 26.01 19639.2 19642.1 9452.2 -0.04 -0.04 2.08 7 32.51 24550.4 24554.3 11856.2 -0.04 -0.03 2.07 8 43.34 32792.3 32797.4 15613.8 0.14 0.14 2.10 9 54.17 40945.1 40953.8 19307.1 0.03 0.04 2.12 10 65.01 49240.3 49247.8 22835.6 0.24 0.25 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3505 to 105441 e- Average count rate measured: 756 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.19 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay