*** MIDAS logfile (05SEPpl1.2) *** Su, 01 Jul 2007 14:06:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 01 Jul 2007 14:07:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070701 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.54 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070701 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1619.1 1619.0 766.5 0.18 0.19 2.11 2 4.34 3223.0 3222.0 1539.3 -0.27 -0.28 2.09 3 8.67 6458.0 6458.3 3080.4 -0.06 -0.05 2.10 4 13.00 9678.1 9676.9 4714.4 -0.16 -0.16 2.05 5 19.51 14529.2 14526.8 6946.1 -0.08 -0.08 2.09 6 26.01 19378.9 19376.2 9365.2 -0.04 -0.05 2.07 7 32.51 24245.8 24242.6 11483.1 0.05 0.04 2.11 8 43.34 32329.8 32328.8 15748.8 0.06 0.06 2.05 9 54.17 40466.8 40467.4 19369.0 0.19 0.20 2.09 10 65.01 48523.8 48524.5 22561.6 0.12 0.13 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3433 to 102900 e- Average count rate measured: 746 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.15 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.28 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay