*** MIDAS logfile (05SEPpl1.2) *** Th, 21 Jun 2007 17:06:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 21 Jun 2007 17:06:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070621 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.69 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070621 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1653.0 1652.9 798.9 0.13 0.12 2.07 2 4.34 3295.8 3295.1 1582.8 -0.18 -0.19 2.08 3 8.67 6595.7 6596.2 3115.4 -0.11 -0.11 2.12 4 13.00 9893.5 9892.9 4735.7 -0.12 -0.13 2.09 5 19.50 14845.3 14847.2 7149.2 -0.09 -0.08 2.08 6 26.01 19805.2 19806.4 9666.1 -0.03 -0.03 2.05 7 32.51 24765.8 24766.7 11980.7 0.00 0.01 2.07 8 43.34 33056.0 33055.5 15870.7 0.11 0.11 2.08 9 54.17 41322.4 41323.8 19593.5 0.12 0.12 2.11 10 65.01 49615.0 49615.8 23282.7 0.17 0.17 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3502 to 105122 e- Average count rate measured: 762 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.18 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay