*** MIDAS logfile (05SEPpl1.2) *** Fr, 15 Jun 2007 15:42:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 15 Jun 2007 15:42:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070615 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.22 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070615 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1645.3 1644.8 796.9 0.12 0.14 2.06 2 4.34 3281.3 3280.8 1553.5 -0.17 -0.17 2.11 3 8.67 6566.4 6566.4 3128.9 -0.10 -0.10 2.10 4 13.00 9848.1 9849.7 4729.0 -0.12 -0.11 2.08 5 19.51 14778.0 14778.5 7141.5 -0.09 -0.09 2.07 6 26.01 19712.2 19712.3 9499.3 -0.04 -0.05 2.08 7 32.51 24649.8 24651.6 11902.7 -0.01 -0.01 2.07 8 43.34 32907.2 32908.3 15672.3 0.12 0.11 2.10 9 54.17 41131.5 41135.0 19689.8 0.11 0.11 2.09 10 65.01 49392.1 49392.9 23050.3 0.18 0.17 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3491 to 104812 e- Average count rate measured: 759 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.18 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay