*** MIDAS logfile (05SEPpl1.2) *** Th, 07 Jun 2007 15:13:27 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 07 Jun 2007 15:13:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070607 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.81 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070607 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 7.9 8.0 32.9 -0.59 -0.02 0.24 2 4.34 15.6 15.5 12.7 1.33 0.25 1.23 3 8.67 30.3 30.3 21.0 -0.08 -0.35 1.45 4 13.00 45.2 45.2 28.1 -0.28 -0.13 1.60 5 19.51 67.5 67.6 38.8 -0.25 -0.10 1.74 6 26.01 89.8 89.8 50.1 -0.30 -0.26 1.79 7 32.51 112.3 112.3 60.2 -0.13 -0.06 1.87 8 43.34 150.1 150.1 81.6 0.21 0.32 1.84 9 54.17 187.2 187.3 96.9 0.08 0.26 1.93 10 65.01 224.4 224.3 114.1 0.01 0.10 1.97 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 18 to 501 e- Average count rate measured: 3 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.23 +/- 0.16 e-/ADU RON = 4.37 +/- 0.31 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.155 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.82/-0.47 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay