*** MIDAS logfile (05SEPpl1.2) *** Tu, 22 May 2007 16:01:47 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Tu, 22 May 2007 16:02:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070522 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.31 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070522 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1672.8 1672.9 811.1 0.15 0.12 2.06 2 4.34 3336.2 3336.0 1583.0 -0.17 -0.19 2.11 3 8.67 6677.4 6677.5 3188.8 -0.11 -0.11 2.09 4 13.00 10015.7 10015.9 4838.1 -0.12 -0.12 2.07 5 19.51 15030.4 15030.6 7210.8 -0.08 -0.08 2.08 6 26.01 20049.7 20049.8 9729.8 -0.04 -0.03 2.06 7 32.51 25071.4 25073.9 12034.6 -0.00 0.01 2.08 8 43.34 33463.6 33465.7 16258.9 0.10 0.11 2.06 9 54.17 41828.3 41834.1 19981.4 0.10 0.12 2.09 10 65.01 50229.1 50227.9 23291.9 0.17 0.17 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3555 to 106742 e- Average count rate measured: 771 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.17 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay